Skip to Content
Rating
0 0

There are no comments for now.

5. In comparison with lower voltage radiographs, high energy radiographs show:
10. Cobalt-59 becomes Cobalt-60 in a nuclear reactor when it captures:
11. In an X-ray tube, the filament and focusing cup are the two essential parts of the:
12. A 250 kVp X-ray machine used with lead foil screens has an approximately practical thickness limit of:
14. Side scatter can cause which type of effect on the radiograph?
16. Sharp, black, bird-foot shaped marks (not corresponding to discontinuities) are probably caused by:
17. A voltage selector consisting of an iron-core transformer with a single tapped winding is called:
18. Decrease in gamma ray energy when passing through a specimen is commonly caused by:
19. The range of specimen thickness that can be adequately recorded on a radiograph is referred to as the:
22. An X-ray tube with a small focal spot is considered better than one with a large focal spot when it is desired to obtain:
24. A device which is basically a combination of magnet and transformer designed to guide and accelerate electrons in a circular orbit to very high energies is called:
25. The maximum practical speed of scanning a test object for conventional fluoroscopic inspection has been estimated to be:
26. Which of the following will NOT materially influence the image density of a radiograph?
31. Constant agitation of the developer during processing by mechanical stirrers or circulating pumps may:
32. Radiographic undercutting is caused by:
34. Films kept between lead screens in a cassette for a long period in high temperature/humidity may:
35. A curve that relates density with the logarithm of exposure or relative exposure is called:
37. When X-rays, gamma rays, light or electrons strike the photographic emulsion, change in silver halide crystals is referred to as:
38. When a faster speed film is substituted for a slower one to improve exposure economics, which condition occurs?
39. Which technique variable is most commonly used to adjust subject contrast?
40. Low voltage X-ray tubes are generally fitted with windows made of:
43. If a processed film allows 1/10 of incident light to pass through, the optical density of the film is:
45. The maximum size of irrelevant indications for ½″ thick welded plate radiographed as per UW-51 of ASME Section VIII Div. 1 will be:
46. The amount of unsharpness of image (Ug) permitted for a 38 mm thick object is:
47. Acceptable density variation in area of interest compared to density through the IQI image:
50. Per ASME Section VIII Division 1, a non-relevant indication is subject to:
52. The thickest wire in ASTM wire-type IQI is in:
53. When using gamma rays, the transmitted minimum film density required on a radiograph is:
57. Periodic calibration verification of densitometer should be performed:
58. For curved component with SFD greater than radius of curvature, location markers are on:
62. To radiograph a 25 mm CS butt weld with Ir-192 source, a typical SFD that gives Ug ≤ 0.020″ and reasonable exposure for a 4 mm × 2 mm source is approximately:
63. A radiograph of a 12 mm CS weld shows IQI no.15 essential hole 2T visible, density 2.2 H&D, no backscatter shadow. The radiograph quality is:
64. Through-IQI density on a radiograph reads 2.50 H&D. Adjacent area-of-interest density reads 3.50. Per ASME V T-282.2:
65. During panoramic exposure of a 6″ NPS pipe weld, you must place the source at the centre. Locating the source 2 mm off-centre with 6 mm wall thickness will:
66. On a stainless-steel weld with Ir-192, you obtain density 1.9 H&D using SS step-wedge IQI #25 (no shim). Per ASME V you should:
67. A welder cap height adds 1.5 mm reinforcement on each side of a 25 mm CS plate weld. What thickness is used to select the IQI?
68. You exposed a film and read step-wedge density 0.05 above the calibrated value. The likely cause is:
69. During interpretation of a CS weld radiograph you observe a dark, sharply rounded indication 1.6 mm diameter. Per UW-51 of ASME Section VIII Div 1 acceptance criteria for ½″ plate, this is:
70. Two parallel linear indications, each 5 mm long and separated by 6 mm, in a 25 mm CS weld. Per UW-51 these are:
74. Density variation across film width on a 12 m circumferential exposure exceeds tolerance in the over-exposed end. The MOST LIKELY cause is:
75. To increase subject contrast on an aluminium part, the technician should:
76. During development the technician notices the radiograph emerging too 'flat' (low contrast). Likely cause:
78. During Ir-192 exposure, your survey meter reads 1 mR/hr at the rope barrier 12 m away. If you halve the distance to 6 m, the reading will be approximately:
79. Lead screen front 0.005″ / back 0.010″ used with X-ray ≤ 250 kVp serves to:
80. To document an exposure, the radiographer's technique sheet should AT MINIMUM include: